Conclusion

330.

It follows from the above that:

a.

Medpro was in breach of contract in that the Gowns did not comply with the requirement of (i) validated process demonstrating sterility level 10-6 (ii) compliance with EN 556-1 (iii) CE marking including notified body number;

b.

DHSC did not effectively reject the Gowns;

c.

It can nonetheless recover the full value of the Gowns as damages: the evidence as to alternative uses of these (effectively) non-sterile Gowns does not show that it was probably possible for them to be sold elsewhere to mitigate the loss;

d.

However DHSC cannot recover the £8 million which it claimed for storage costs. That loss was not proved on the evidence adduced at trial;

e.

Medpro has no counterclaim against DHSC either for rectification or damages.